Installation
IMPORTANT
18
Operating Instructions BioPAT
4.4

Connection

Making connections while the BioPAT
still connected to power or with the wrong kind
of voltage can cause damage!
Use only cables from the equipment supplied
t
with the BioPAT
equipment. The module is not equipped with
an ON | OFF switch. First, leave the control
unit and | or supply unit of the bioreactors
switched off and | or disconnect the external
power adapter from the mains current. First
connect the cables for the signal transmission
and power supply.
Connect the exhaust line from the Biostat
t
and B-DCU II.
Make sure that the lines of the culture
t
vessel contain no condensation.
Connect the exhaust line to the BioPAT
t
Xgas.
If exhaust air entering the room air can be
t
regarded as safe, you can leave the
BioPAT
Xgas outlet open. If the exhaust
®
air contains gases in potentially hazardous
concentrations, connect the outlet to the
exhaust air fittings in the laboratory.
Xgas
®
Xgas and | or the bioreactor
®
Xgas is
®
B
®
®