10.4.3
Scattered Dose
Scope
IEC 60601-2-65 203.13
Requirements
ME EQUIPMENT shall be provided with means to optionally allow actuation of the
EXPOSURE from a PROTECTED AREA after installation.
Relevant instructions shall be given in the ACCOMPANYING DOCUMENTS.
Results
The following exposure timetables were established with a unit equipped with a cone
that corresponds to a focus-to-skin distance of 200 mm (8 inches) respectively.
Instrument
X / Gamma
Survey Meter
EzRay Air Portable (Model: VEX-P300) User Manual
Test Equipment
Manufacturer
Radcal Co.
PMMA Phantom aligned to 280 mm away from Focal Spot
Max. Exposure Condition
Measure point: 1000 mm from PMMA Phantom
Model
9015/10X5-180
Method
10. Appendix
S/N
91-1470/19069
79