Werbung

5. Service Mode
5.3.3
User Default Settings (for each patient and tooth type)
Perform each mode to set the default settings for a specific patient and a tooth type
referring to the table below.
Service Mode No.
n.10
n.11
n.12
n.13
n.14
n.15
To set the default setting for a specific patient type and a tooth type,
1.
Select the Service Mode number between n.10 (Adult) and n.11 (Child) for a
specific patient type, and from n.12 to n.15 for a tooth type. For instance, when
select n.10 by pressing the jog dial to set the patient type 'adult' as default, then
"YES" is displayed on the Control Panel.
2.
Save the setting, and check if the patient type 'adult' has been saved as default.
3.
When performing other settings, follow the same procedure.
46
Item
Adult
Child
Incisor
Canine
Molar/Premolar
Bitewing
EzRay Air Portable (Model: VEX-P300) User Manual