5. Service Mode
5.3.5
Perform this mode to set the angle increments when using the Auto Mode.
Service Mode No.
n.18
When using the Auto Mode, the angle value increases and decreases according to the
increments setting (default: 1 degree).
The angle increments can be set from 1 to 5 degrees.
To change the angle increments setting,
1.
Select n.18 by pressing the jog dial, then the default angle increments "001°" is
displayed on the Control Panel as shown in the following figure.
2.
Turn the jog dial to adjust the angle increments you want, and save the setting.
3.
Check if it has been correctly saved.
48
Angle Increments Setting
Item
Angle increments setting