2. Safety Instructions
Radiation Safety
10
If the device not in use has been turned on for long periods of
time, the battery may be fully discharged.
- Depending upon the battery discharge status, it takes
about 1 day for charging the battery. If the device is not
turned on after charging the battery for about 1 day, it
indicates that the battery has been fully discharged.
Contact your Service Representative for battery
replacement.
- DO NOT charge a fully discharged battery, as this may
cause fire or explosion. Be sure to replace the battery
(provided by VATECH).
Users can replace batteries.
When charging the battery, the exposure function is locked.
Be sure to turn off the device when not in use. This helps
to ensure the life of the battery.
Be sure to charge the battery frequently. This helps to
ensure the life of the battery.
When using the device, it is recommended that all users
comply with the following radiation safety guidelines for the
safety of the users and the patients.
All users and patients should wear a protective device,
such as a lead apron, thyroid collar, etc.
This device should be operated in the area that is more
than 6 feet away from other personnel, such as assistants
or other patients. If they should stay closer than 6 feet, it is
recommended that they wear a lead apron, thyroid collar,
or stay behind a lead shield.
Pregnant women should not be exposed to X-rays unless it
is strictly necessary.
All users should comply with the Radiation Protection
Policies established by the government.
When selecting a Position Indicating Device, it should be
considered if the PID can be used with the backscatter
shield attached at the outer end of the cone for most
operator protection.
EzRay Air Portable (Model: VEX-P300) User Manual