6. Troubleshooting
Troubleshooting
Problem
The device is not
turned on.
Control Panel is not
turned on.
No X-ray emission
X-ray emission
works, but
exposure is too
light or
completely white.
X-ray emission
works, but
exposure is too
dark.
56
Cause
The power switch is not
turned on properly.
Battery discharged
Battery cable is not
correctly connected.
Defective battery
Defective mainboard
Internal cable
disconnected
The generator is cooling.
Defective Remote
Exposure Switch
Internal cable
disconnected
Defective generator
Tube lifecycle termination
The device has been
positioned incorrectly.
Exposure time is too
long.
The receptor is facing the
wrong way.
Exposure time is too
short.
EzRay Air Portable (Model: VEX-P300) User Manual
Solution
Turn the device power switch off
and turn it back on.
Recheck after charging the battery
with a charger.
Contact your Service
Representative.
Contact your Service
Representative.
Contact your Service
Representative.
Contact your Service
Representative.
Wait for the cooling time (refer to
'Duty Cycle').
Contact your Service
Representative.
Contact your Service
Representative.
Contact your Service
Representative.
Contact your Service
Representative.
Adjust the position of the device.
Decrease the exposure time.
Reposition the receptor.
Increase the exposure time.