Herunterladen Inhalt Inhalt Diese Seite drucken

VATECH EzRay Air Portable VEX-P300 Bedienungsanleitung Seite 65

Vorschau ausblenden Andere Handbücher für EzRay Air Portable VEX-P300:
Inhaltsverzeichnis

Werbung

7.2.1
Maintenance Task Checklist
Before the operation, ensure that the device is clean and ready
for use.
After using the device, make sure that the device has been
turned off.
Wipe the outer covers of the device with a dry cloth at the end of
each day's operation.
Ensure that the signal is audible, and the X-ray emission light is
visible when making an exposure.
Ensure that the yellow (exposure) indicator light turns on when
the Exposure Button is pressed.
Ensure that the battery charging LED indicator comes on when
charging the battery.
Ensure that the battery level indicator displays at least two levels
(Battery Level 2). For more information on the battery levels, see
'4.5 Using the Battery'.
Ensure that all visible labels are intact and legible.
EzRay Air Portable (Model: VEX-P300) User Manual
Always turn off the device before performing any maintenance.
Tasks
DO NOT use detergents or solvents to
clean the outer covers of the device.
If any defects are found, do not operate the device since it has
to be handled by a qualified person. Contact your Service
Representative.
7. Cleaning and Maintenance
Period
Daily
Daily
Daily
Daily
Daily
Daily
Daily
Monthly
59

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis