MACS II
Mechanical alkaline cleaning and thermal disinfection
Machine type: Single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Prerinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfec-
tion
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at
least)
Mechanical cleaning/disinfection with man-
ual pre-cleaning
Note
The disinfector must be of tested and approved effec-
tiveness (e.g. DGHM or FDA approval or CE mark
according to DIN EN ISO 15883).
Note
For thermal disinfection, always use fully desalinated
water (demineralized, low microbial contamination:
drinking water quality at least) and ensure that Ao is
>3 000 for the process.
Note
The disinfector used for processing must be serviced
and checked at regular intervals.
30
T
t
Water
[°C/°F]
[min]
quality
<25/77
3
D–W
55/131
10
FD-W
>10/50
1
FD-W
90/194
5
FD-W
-
-
-
Chemical/Note
-
BBRAUN HELIMATIC CLEANER alcaline
with tensides, application solution 0.5%
-
-
According to disinfector program