Herunterladen Inhalt Inhalt Diese Seite drucken

Timer Mode Unimax 1010, Promax 1020 - Heidolph Hei-MIX Unimax 1010 Betriebsanleitung

Temperature controlled shaking
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Operation

Timer mode Unimax 1010, Promax 1020

In timer operation, a specific period of time can be set for continuous shaking.
Place the vessel(s) on the platform. Observe the instructions in section
"Attachments" on page 41.
Ensure that the speed controller is set to zero.
Switch on the device (see "Switch the device on/off" on page 42).
Use the two timer buttons to set a value in the range of 0 – 999 minutes (LED
Time
flashes).
Press and release the buttons to increase or decrease the display
value in one-step increments.
Press and hold the buttons to increase or decrease the display value
in fast forward.
The set timer value is retained until the device is switched off and
can be adjusted at any time during operation.
Press the start/stop button to activate the motion mode.
start/stop
The
The the time already elapsed is displayed on the device.
Adjust the shaking intensity with the speed control as required.
The shaking intensity (speed) is shown in the display.
In case of stopping the movement by pressing the start/stop
button in timer mode, the timer is stopped but not reset. The
timer continues to run accordingly when the shaking movement is
restarted, starting from the last count value.
In the event of a power interruption or when the device is switched
off via the main switch, the timer is reset to zero!
Use the Speed/Time mode selector to toggle between the timer and the speed
display (LED Time/Speed) while the system is running.
Press Stop/stop button again to deactivate the motion mode.
Time
and
LEDs light up.
45

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis