Operation
 Press the ZERO/MENU button.
appears in the display for 2 s. The ZERO function can
2Ero on
be activated.
 Press the SET/OK button to perform the zero point correction.
 The display and the MIN/MAX values are reset.
or
appears in the display for 2 s. The measured pres-
oFL 2Ero
sure (0 bar) is greater than 5% of the pressure range. The ZERO
function cannot be performed.
 The device switches back to measuring mode.
 Depressurise and press the ZERO/MENU button again.
Resetting the zero point correction
The zero point correction stays activated until the device is switched
off. After it is switched on again, the zero point correction is no longer activated.

5.2 Device settings

Automatic shut-off
(
/
Po off
Po on
Measuring unit (
Display filter (
Filt
Delete all measured data**
(
dEL dAtA
Device information
* Value selection: 0...7 (0 = no damping)
** Only version with data logger
You will automatically return to the display mode if you do not press any button for about 5 s.
Automatic shut-off
When the function is activated, the device automatically switches off after 5 minutes.
The automatic shut-off is deactivated for devices with data logger during data recording.
If the function is deactivated, the device is in continuous duty and has to be manually
switched off with the ON/OFF button.
The current setting is displayed when the digital pressure
gauge is switched on:
= activated (automatic shut-off);
Po on
= deactivated (continuous duty).
Po off
The settings
Po on
tive again when the device is switched on.
- 28 -
Press the ZERO/MENU button for 2 s.
Press ZERO/MENU button to continue.
)
)
unit
Press the ZERO/MENU
)*
button for the next menu item.
)
or
remain stored and are ac-
Po off
Confirm your selection with
the SET/REC button. You
return to the display mode.