Device Description - Sika E2 Betriebsanleitung

Digitalmanometer
Vorschau ausblenden Andere Handbücher für E2:
Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

E2 / D2 / C2
2

Device description

The types E2 / D2 / C2 are digital pressure gauges with actual value and MIN/MAX display.
They have an accuracy of ±0.5% Type E2, ±0.1% Type D2 and ±0.05% Type C2, relative to
the corresponding pressure range end value (FS).
Signals from the pressure measurement cells are recorded at a sampling rate of 10 ms
(100 measurements/s), converted into pressure values and displayed. Thanks to the high
sampling rate, dynamic pressure peaks can also be measured. They are written to the
MIN/MAX memory, which is continuously updated.
Versions
Standard
Accuracy (from end value)
Pressure range
-1...3 bar
-1...5 bar
-1...10 bar
-1...20 bar
-1...40 bar
-1...60 bar
0...100 bar
0...160 bar
0...250 bar
0...400 bar
0...700 bar
0...1000 bar
Technical changes reserved
Resolution
1 mbar
1 mbar
1 mbar
1 mbar
10 mbar
10 mbar
10 mbar
10 mbar
100 mbar
100 mbar
100 mbar
100 mbar
Version with data logger
E2 0.5%
D2 0.1%
Device description
C2 0.05%
- 23 -