4.2.6 Calibration Test
Calibration test procedures:
1.
Long press the Menu button until you see
2.
Short press the No button several times to navigate until you see
3.
The display flashes with
menu. With no weight on the pan, press the Yes button to establish the zero point.
4.
The display shows
5.
The display flashes with the calibration weight and the unit of the last time. (For example,
6.
To change the test calibration weight:
 i-DT61PW: short press the No button several times until the desired digit appears. Short press the Yes
button to accept the digit and move to the next one. Repeat the process until all the digits are correct.
Press the Yes button to accept the calibration point.
 i-DT61XWE: press the numeric keys to edit the weight. (Do not press the Yes button here until you
finish step 7).
7.
Place the specified test weight on the pan and press the Yes button.
8.
The display flashes with the difference between the calibration data and the test weight. (For example,
0.010
kg). If the terminal is connected to a printer or other devices, the result of the Calibration Test will be
printed.
9.
After 5 seconds, the test ends and the scale returns to the active weighing mode with the display of the
current weight.

4.2.7 End Cal [End]

End
When
is displayed, press the Yes button to exit this menu and advance to the next Sub-menu or press the
No button to advance to the first menu item in the this Sub-menu.
[
]
C.test
0
and the calibration unit based on the capacity and unit you set in the capacity
--t--
while the zero point is recorded.
C.A.L
. Press the Yes button.
Defender 6000 Indicators
C.test
. Press the Yes button.
015.000
kg).