EN-4

2.5 Initial Calibration

When the scale is first installed, it should be calibrated to ensure accurate weighing
results.
Press and hold the ENTER/Cal button until CAL is displayed. When the button is
again pressed and released, the display shows –C- while zero reading is stored.
Next, the display shows the calibration weight value in grams. Place the specified
calibration weight (sold separately) on the pan. See Table 2-1 for appropriate
weights. The display shows –C- while the reading is stored. The display returns to
the previous application mode and scale is ready for use.
V31XH202
Calibration
200g
Weights
3. OPERATION

3.1 Weighing Mode

Use this mode to measure the weight of a sample in the selected unit of measure.
1. Press MODE/UNIT/Menu until the desired unit icon is displayed.
2. Put an empty container on the pan, then press O/T/On/Off to tare the container
weight.
3. Add material to the container. The display shows the weight of the material.
Figure 2-3. Power Connection and Sealing
TABLE 2-1. CALIBRATION WEIGHTS.
V31XH2
V31X3
2 kg
3 kg
MODELS
V31X6
V31XW301
6 kg
300 g
Valor 3000 Series Scales
.
V31XW3/
V31X3N
3 kg
V31XW6/
V31X6N
6 kg