5.
Install the waste bag onto the waste bag pins.
7.
Place the bowl in the centrifuge chamber with the yellow-
capped waste line outlet tubing facing the waste bag.
Refer to the markings on the top panel.
Wash kit setup
6.
Remove the holding bag from the wash kit and hang it on
an upper IV pole hook. Close the clamps on the 2 red-
capped outlet lines that each include a membrane port.
8.
Align the centrifuge markings with the vertical assemblies
on the bowl. To lock the bowl, simultaneously press down
on the bowl and turn it clockwise until a click is heard.
English
5