If necessary, adjust the vacuum setting using the Vacuum
8.
drop-down menu on the Main screen.
10.1. Before blood collection, open the roller clamp and prime
the collection reservoir with a minimum of 100 mL of
anticoagulant solution.
Blood collection system setup
9.
Spike the anticoagulant container. If the container is
nonvented, open the vent cover on the drip chamber.
10.2. Reduce the anticoagulant flow to a ratio of approximately
15 mL of solution to 100 mL of blood.