Prior to use, check that the product is in good working order.
►
Follow TA022450 "Notes on electromagnetic compatibility (EMC) for
Acculan
components",
https://extranet.bbraun.com
►
To prevent damage caused by improper setup or operation, and in order
not to compromise warranty and manufacturer liability:
– Use the product only according to these instructions for use.
– Follow the safety and maintenance instructions.
– Only combine Aesculap products with each other.
►
Ensure that the product and its accessories are operated and used only
by persons with the requisite training, knowledge, or experience.
►
Keep the instructions for use accessible for the user.
►
Always adhere to applicable standards.
►
Ensure that the electrical installation of the room is consistent with
the requirements of IEC/DIN EN.
►
Do not operate the product in explosion-hazard areas.
►
Sterilize product before use.
►
When using the ECCOS holder system, adhere to the appropriate
instructions for use for TA009721, see Aesculap Extranet at
https://extranet.bbraun.com
4.
Product description
4.1
Scope of supply
Art. no.
Designation
GA344
Small drill
GA344244
Kirschner wire protection sleeve
GA678
Sterile funnel
GA344211
Rinsing adapter
GB600880
Oil spray adapter for GA344/GA844
TA011944
Cleaning brush
TA014551
Instructions for use for GA344 (flyer)
4.2
Components required for operation
■
Li-Ion battery short GA346 (charged)
■
Sterile funnel GA678
■
Cap GA675
■
Attachment (depending on indication)
■
Tool (depending on indication)
see
Aesculap
Extranet
4.3
Operating principle
The product 1 contains an electric motor, which is powered by a replace-
at
able battery 7.
The charged non-sterile battery 7 is inserted into the product 1 via the
sterile funnel 6 and sealed with the cap 8 so that it is sterile.
The output speed is controlled electronically and can be continuously reg-
ulated with the knob for speed control 2.
The direction of rotation can be changed from right to left again using the
knob for left rotation 11. The product 1 can also be operated in an oscil-
lation mode.
The product 1 has a coupling on the output side, which allows for the
attachment of various drilling, medullary reaming and saw attachments.
These attachments lock automatically when attached to the product.
The attachment can be removed by operating a rotating sleeve 13. Drives
in the various attachments change the motor speed and enable operation
of the motor at an ideal operating point.
The attachments have various integrated couplings on the working end, to
enable attachment of various tools, with adapter if necessary.
The product 1 is equipped with a cannulation to attach guide shafts or the
like.
5.
Preparation
Aesculap assumes no liability if the following rules are not followed:
►
Do not use products from open or damaged sterile packaging.
►
Prior to use, inspect the product and its accessories for any visible
damage.
►
Use the products and their accessories only if they are in perfect tech-
nical condition.
6.
Working with the device
Risk of infection and contamination!
Product is delivered unsterilized!
WARNING
►
Sterilize the product before use according to the
operating instructions.
Risk of injury and material damage due to acciden-
tal activation of the product!
►
Products which are not being actively used must
WARNING
be secured against accidental activation (posi-
tion OFF).