Aesculap
Small drill GA344
7.7.2
Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection
V
Drying
D-W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
*Recommended: BBraun Helimatic Cleaner alkaline
►
Insert product 1 in the correct position in the ECCOS holder, see Fig. D.
►
Attach the rinsing adapter 17 (GA344211) to the product 1 in the
mount 14.
►
Then connect the interior rinse device and connect to the rinse connec-
tor of the cleaning/disinfection device/rinse cart.
►
Attach the Kirschner wire protection sleeve 15 to a rinse hose.
►
After automatic cleaning/disinfection, check visible surfaces for resi-
dues and repeat the cleaning/disinfection process as needed.
7.8
Inspection, maintenance and checks
►
Allow the product to cool down to room temperature.
►
Spray product after every cleaning and disinfection with oil spray
adapter 18 GB600880 (green) approx. 2 sec with Aesculap-STERILIT-
Power-Systems oil spray GB600, see Fig. C.
Note
Aesculap additionally recommends occasional spraying of moving parts
(such as knobs, couplings, cover lids) with the Aesculap STERILIT-Power-
Systems oil spray.
►
Check the product after each cleaning and disinfection for the follow-
ing: cleanliness, damage, function, abnormal operation noise, exces-
sive heat or heavy vibration.
►
Set aside the product if it is damaged.
12
Acculan 4
®
T
t
Water
[°C/°F]
[min]
quality
<25/77
3
D–W
55/131
10
FD–W
>10/50
1
FD–W
90/194
5
FD–W
-
-
-
Chemistry/Note
-
■
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
■
0.5 % working solution
– pH ~ 11*
-
-
mind. 10 min at max. 120 °C
7.9
Packaging
►
Always follow the instructions for use of the respective packaging and
storage devices (e.g. instructions for use TA009721 for Aesculap-
ECCOS holder system).
►
Insert the products in the correct position into the ECCOS holders, see
Fig. D.
►
Pack trays appropriately for the sterilization process (e.g. in Aesculap
sterile containers).
►
Ensure that the packaging will prevent a recontamination of the prod-
uct.