Herunterladen Inhalt Inhalt Diese Seite drucken

Troubleshooting List - B. Braun Aesculap Acculan 4 Gebrauchsanweisung/Technische Beschreibung

Kleine bohrmaschine
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 20
Aesculap
Small drill GA344
9.

Troubleshooting list

Have defective products repaired by Aesculap Technical Service, see Technical Service.
Issue
Product not running
Product becomes too hot Excessive use
Insufficient power
Loud running noise
14
Acculan 4
®
Cause
No battery
Battery not charged
Battery defective
Product in locked OFF position Safety catch is in OFF position
Product defective
Reprocessing/maintenance
carried out incorrectly
Fall damage, product defec-
tive
Use of a defective attachment Product heating
Blunt tool
Product defective
Product is operating in left
rotation mode
Use of a defective attachment Attachment very hot
Blunt tool
Defective product drive/ball
bearings
Defective attachment
drive/ball bearings
Detection
No battery in the battery shaft
No signal tone when inserting the bat-
tery
No signal tone when inserting the bat-
tery
Product not running
Product heating
Product heating
Product heating
Heating of tool and product
Insufficient product performance
Large amount of heat generated after a
short time
Serrated tool runs in counterclockwise
mode
Tool blades worn out
Loud, perceptible noise during opera-
tion
Loud, perceptible noise during opera-
tion
Remedy
Insert battery.
Charge battery in charger.
Have the manufacturer repair the battery.
Turn the safety catch into the ON position.
Have the manufacturer repair the product.
Follow operating instructions (operating
mode).
Follow operating instructions (preparation,
care).
Preventative: lubricate product before every
sterilization.
Have the manufacturer repair the product.
Have the manufacturer repair the attachment.
Change tool.
Follow operating instructions (preparation,
care).
Preventative: lubricate product before every
sterilization.
Have the manufacturer repair the product.
Follow operating instructions (operating
mode).
Have the manufacturer repair the product.
Operate serrated tool in right rotation.
Have the manufacturer repair the attachment.
Change tool.
Follow operating instructions (preparation,
care).
Preventative: lubricate product before every
sterilization.
Have the manufacturer repair the product.
Follow operating instructions (preparation,
care).
Preventative: lubricate attachment before
every sterilization.
Have the attachment repaired by the manu-
facturer.

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Ga344

Inhaltsverzeichnis