Setting the capture range
The capture range allows unwanted zero deviations to be suppressed. The
setting affects the measured value which is output.
This feature is used to define the capture range and the capture value limits.
Setting the calibration interval
The calibration interval defines the time window for the next calibration. If
the interval expires, the gas detector issues a warning.
Sensor self-test
The sensor self-test is a safety feature that is performed at recurring
intervals. If the sensor does not pass a sensor self-test, a warning is output.
Further negative self-tests will result in an error message being output.
The sensor self-test can be disabled in PolySoft or carried out manually.
For safety-related applications, the sensor self-test must be activated.
Gas settings
Depending on the sensor, the gas type, measuring range and measuring
units can be set with this feature.
14.2
Resetting the gas detector to the factory
settings
Can be performed from 10 minutes after turning on the gas detector.
Device settings, data for WirelessHART
®
configuration, and Bluetooth
pairing PIN are reset.
The feature is also available in PolySoft.
Procedure using the magnetic pen
Placement of the magnetic pen
0 s
Reset the gas detector to the factory settings:
1. Hold the magnetic pen on the mark (Fig. A 12) for approx. 15 seconds until
the blue LED flashes rapidly.
2. Lower the magnetic pen for 7 seconds.
Instructions for use
|
Dräger Polytron
®
network integration, sensor
LED flashes
15 s
0 s
7 s
8 s
®
6100 EC WL, Dräger Polytron
3. Put the magnetic pen back on the marking for 1 second (LED stops flashing)
and then take it off.
✓ The gas detector restarts and is reset.
14.3

Changeable safety-relevant settings

PolySoft
Factory setting
000000
®
Bluetooth
-pairing PIN
Measured gas
sensor-specific
Sample gas module
sensor-specific
Measuring range
sensor-specific
Upper capture value limit
sensor-specific
Lower capture value limit
sensor-specific
Test gas
sensor-specific
Test gas unit
sensor-specific
Test gas concentration
sensor-specific
Calibration interval
sensor-specific
A1 alarm
sensor-specific
A2 alarm
sensor-specific
Capture value
sensor-specific
Measured gas LEL factor
sensor-specific
Serial number of the gas detector
®
WirelessHART
Tag Name
®
1
WirelessHART
Network ID
®
Repeater WirelessHART
|
Settings
®
en
61