Commissioning
1. Turn on the power supply or install the battery box (see "Installing the
battery box", page 46).
 Start sequence begins.
Green LED flashes.
 LED test in progress.
Yellow and red LEDs flash (repeater: Yellow only).
 Warm-up phase 1 starts.
Device variable status is "Bad" and yellow LED flashes, blue LED
®
flashes (Bluetooth
is available).
 The start sequence is complete and the gas detector is ready for
operation.
 Warning signal is output.
Green and yellow LEDs flash.
2. Integrate the gas detector into the network (see "Network integration",
page 47).
3. Wait for warm-up phase 1 and 2.
The duration of the warm-up phase depends on the sensor used (see
sensor data sheet).
Settings with PolySoftcan already be made.
 At the end of warm-up phase 1, the measured value is available, the
device variable status is "Good" and the yellow LED is off.
4. Calibrate the gas detector (see "Calibration", page 50) as soon as warm-up
phase 2 is completed.
5. Test the signal chain by applying target gas to the sensor.
8

Network integration

8.1
Network integration
Also pay attention to the documentation of the Access Point and System
Manager used for network integration.
Network ID and JoinKey must be set in the system manager and gas detector.
Afterwards, the gas detector needs to be restarted.
Instructions for use
|
Dräger Polytron
®
6100 EC WL, Dräger Polytron
Sensor information is necessary to integrate burst mode (Burst Mode).
Network integration should be done with the correct sensor inserted in the
device.
Dräger recommends activating the sensor lock feature after commissioning
the device.
8.2
Connecting to WirelessHART
®
1. Connect device via Bluetooth
2. Enter network ID and JoinKey on "WirelessHART
connect to network.
8.3
Signal transmission to the controller
Recommended update period (Update Period): 1x every 8 seconds
®
(WirelessHART
-Gateway).
Measured value updates (update period, maximum update period
(Maximum Update Period) and timeouts can be set via the WirelessHART
gateway.
9

Burst Mode

WARNING
Mortal danger!
Burst mode configurations (Burst Mode Configurations) have a significant
influence on transmitting the gas reading to the gateway.
Incorrect configuration will result in critical gas concentrations not being
detected or being detected with a long time delay.
► Check parameters at the gateway.
®
Repeater WirelessHART
Commissioning
®
network
to PolySoft.
®
configuration" tab and
®
|
en
®
47