Command 48: Reading additional device status

Byte
Bit
Description
0
0
VOLTAGE_OUT_OF_RANGE
2
BATTERY_WARNING_LOW
6
SENSOR_WARMUP
1
1
SENSOR_VITALITY_FAIR
2
SENSOR_VITALITY_POOR
4
CALIBRATION_INTERVAL_EXCEEDED
7
DEVICE_TEMPERATURE_WARNING
Instructions for use
|
Dräger Polytron
Extended device status
Description
Device variable value
Device variable status
Meaning of "Quality" (Bit 6 and
7):
3: Good – device variable
value with normal accuracy.
2: Poor – device variable
value with low accuracy.
1: Manual/Fixed – device vari-
able value fixed.
0: Bad – There may be a
device variable value, but the
device variable value may not
be used.
®
6100 EC WL, Dräger Polytron
Cause
Cause
Battery voltage below warning limit.
Sensor in warm-up phase 2. No calibration
possible.
The residual sensitivity is between 25 and
50 %.
The residual sensitivity is between 0 and
25 %.
Calibration interval expired.
NTC sensor greater or smaller than permit-
ted temperature.
®
Repeater WirelessHART
Troubleshooting
Remedy
Remedy
Contact DrägerService.
Change battery pack.
Wait for warm-up phase 2 to finish.
Consider a sensor change.
Consider a sensor change.
Calibrate the sensor.
In case of high temperatures due to solar radi-
ation, use weather protection set (accessory).
®
|
en
53