Cleaning and maintenance
10.3

Replacing the sample tube

7
- Loosen fixing screw (7) of sample tube in the direction of the arrow and re-
move old sample tube.
- Insert new sample tube with washer into screw gland of sample tube.
- Tighten screw gland of sample tube to instrument by hand.
- Check sample tube for secure seating.
10.4

Replacing the sample pump

Removing cover of sample pump
- Slightly press together cover of sample pump on grip recesses and detach it.
- Pull sample pump out of guide system.
Inserting sample pump
- Insert new sample pump into guide system.
2
3
Ensure that the pump plunger (2) and pump body (3) are situated in the corre-
sponding guides when inserting the sample pump.
- Check for proper functioning of sample pump.
- Put the cover of the sample pump back in its place.
22