PC settings

- Configure the infrared interface of the Densito 30P as described in Chapter 5.6.
- Connect the adapter to an available serial interface (COM1, COM2, etc.) on
the PC.
- Insert the accompanying CD-ROM into the CD-ROM drive of the PC.
- Install the MiniTerm program.
- Start the MiniTerm program on the PC.
- Select the serial interface (Port) to which the infrared adapter is connected.
- Configure additional settings according to the figure below.
- Press the OK button on the PC.
The Connect window is closed.
8.3

Data format

If RS is selected as interface, the data is transferred in the following format:
Type
1)
Sample No.
,
Data column
STX
Start byte = 0
1
1
1
Type
Result
Data column
Start byte = 10
7
Type
Temperature
Data column
Start byte = 29
5
α
Type
4)
,
5)
Data column
A
Start byte = 37
1
1
1
1)
Sample identification (a...z, space)
2)
If result is marked (*); otherwise, space
Temperature unit (°C or °F)
3)
Number of the temperature-compensation coefficient
4)
The first 8 bytes contain only spaces, unless Comp. Density, SG, °Baumé or
5)
Conc. is selected as the result unit
20
manuals search engine
2)
,
,
4
1
1
1
Unit
,
1
3)
,
,
1
1
1
α-value
CR
LF EOT
1
1
1
5