Measurement (meas)
6.3

Filling the measuring cell using an external syringe

11
3
6
- Press the drain button (3).
- Slide up the lock button (11).
- Remove the screw stopper (6) with a coin.
- Screw in the adapter for the external syringe.
- Hold the sample tube over a waste container.
- Slowly inject the sample into the measuring cell.
- Ensure that there are no air bubbles in the measuring cell.
- Do not remove the syringe while measuring.
6.4

Measuring

The procedure is dependent upon the settings in the menu, see Chapter 5.4.
The
symbol appears in the display: Measuring with automatic stability
control
- Press the ok/meas. key to start measuring.
The
symbol blinks during measurement. The result appears in reverse dis-
play.
The
symbol does not appear on the display: Measuring with manual stability
control
The instrument measures continuously.
- Wait until the displayed value stabilizes.
- If the result is to be saved: Press the ok/meas. key.
The result appears in reverse display.
Measuring with temperature-compensation coefficients (α)
(Comp.Density, SG, Conc. or Baume on the display)
- Press the
key until a previously-saved temperature-compensation coeffi-
cient appears in the display, e.g.
- Select the desired coefficient using the
- Confirm the selected coefficient with the ok/meas. key.
- Measure using either automatic or manual stability control.
Emptying the measuring cell
- Hold the sample tube over a waste container.
- Press drain button (3) downward slowly and move it up again with the fill
button (12).
14
manuals search engine
α1=0.132
.
and
keys.