To Replace the Filter
(continued)
3. Remove and properly dispose of the used
filter.
4. Mark the replacement date on the label of
the new filter.
5. Install the new filter into the filter
compartment and replace the door
(Figure 2).
6. Make sure the power cord is connected
between the rover and facility power.
7. Push the rover power switch to the ON
position and acknowledge the HIGH
SUCTION DEVICE WARNING.
8. Press the SUCTION ON/OFF button and
turn the SUCTION LIMIT SETTING dial of
either canister to the maximum suction
level. Make sure the maximum suction
level value on the CONTROL PANEL
DISPLAY is obtained (Figure 3).
4
Print Date: Mar 07, 2017 04:15:37 PM
0000012816, Rev. H Effective Date: Mar 7, 2017 12:56:42 PM
NOTE: Make sure there are no manifolds
installed in the rover.
9. If the suction level is NOT reached,
reposition the filter. If the suction level
is still NOT reached, contact the Stryker
Neptune Service Center.
Figure 3 – To Apply Suction
0702-034-700 Rev-H
60
mm-Hg
SETTING
LOW
480
mm-Hg
HIGH
SETTING