General Setup (applies for all parameters)

21.7
%sat
25.0
°C
MENU
Quick Setup
Channel configuration (for multi-channel devices)
6.0
A
pH
25.0
A
°C
a =
pH
Analog Output
Yes
Analog outputs
7.0
pH
25.0
°C
Aout1 min= 0.000 pH
u
Aout1 max= 14.00 pH
Setpoints
7.0
pH
25.0
°C
a Set Point Yes
u
SP1 Type= Off
While in Measurement mode press the [MENU] key to bring up the Menu selection.
Select "Quick Setup" and press the [ENTER] key.
Display Convention:
1st line on display = > a
2nd line on display = > b
Select the units of measurement for a and b. Only lines a and b can be configured in
Quick setup. Go to the Configuration Menu to configure lines c and d.
The user can configure the second channel by changing a to c, using the
Select the units of measurement for a and c. Only lines a and c can be configured in
Quick setup. Go to the Configuration Menu to configure lines b and d.
m
By selecting "Yes" the linear 4–20 mA analog output Aout1 will be set up for measure-
ment a when [ENTER] is pressed. Selecting "No" means that no analog output is
set up.
Aout1 min, Aout1 max are the minimum and maximum measurement values for the
4 and 20 mA values respectively.
Note for multi channel devices: The user can configure output Aout3 to
measurement c by going back to the previous menu and selecting c.
After configuring the Analog Output a setpoint can be configured for that output. If "No"
is selected and [ENTER] is pressed then the quick setup is done and the menus are
exited without setting up any setpoint.
(PATH: Menu/Quick Setup)
3rd line on display = > c
4th line on display = > d
or
key.
m
.
7