General Setup (applies for all parameters)

21.7
%sat
25.0
°C
MENU
Quick Setup

Channel Selection

A
6.0
pH
A
25.0
°C
Channel Select = ISM
Parameter = Auto
Analog outputs
7.0
pH
25.0
°C
Aout1 min= 0.000 pH
Aout1 max= 14.00 pH
While in Measurement mode press the [MENU] key to bring up the Menu selection.
Select "Quick Setup" and press the [ENTER] key.
Display Convention:
1st line on display © a
2nd line on display © b
Select the units of measurement for a and b. Only lines a and b can be configured in
Quick setup. Go to the Configuration Menu to configure lines c and d.
Please select the type of Sensor:
Analog: For conventional analog sensors (will be displayed on channel "A").
ISM:
For ISM sensors (will be displayed on channel "B").
Please select the parameter requirement:
The choice of parameter depends on the level of transmitter. If an ISM sensor is select-
ed, the setting "Auto" means, all possible ISM sensors will be recognized and accept-
ed. If a special parameter is chosen, only this parameter will be recognized and ac-
cepted on the transmitter.
By using analogue 2-e or 4-e conductivity sensors,
see "Conductivity Quick Setup" below for intermediate steps.
By selecting Yes the linear 4–20 mA analog output Aout1 will be set up for measure-
ment a when [ENTER] is pressed. Selecting No means that no analog output is set up.
Aout1 min, Aout1 max are the minimum and maximum measurement values for the
4 and 20 mA values respectively.
Note for multi channel devices: The user can configure output A out 3 to
mesurement c by going back to the previous menu and selecting c.
(PATH: Menu / Quick Setup)
3rd line on display © c
4th line on display © d
9