Fora Diamond MINI Bedienungsanleitung Seite 19

Blutzucker-überwachungsssystem
Vorschau ausblenden Andere Handbücher für Diamond MINI:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 22
Blood Glucose Measurement
Symptom
Cause
Batteries
exhausted.
The device
does not
display a
Test strip inserted
message after
upside down or
inserting a
incompletely.
test strip.
Defective device
or test strips.
Insufficient blood
sample.
Defective test
The test does
strip.
not start after
Sample applied
applying the
after the device
sample.
is automatically
turned off.
Defective device.
What To Do
Recharge the battery
immediately and reset
date and time on the
meter setting.
Insert the test strip with
contact bars end first and
facing up.
Please contact customer
service.
Repeat the test using a
new test strip with larger
volume of blood sample.
Repeat the test with a new
test strip.
Repeat the test with a
new test strip. Apply
sample only when
flashing " " appears on
the display.
Please contact customer
service.
Error in
performing the
test.
Control solution
vial was poorly
shaken.
Expired or
The control
contaminated
solution
control solution.
testing result
is out of
Control solution
range.
that is too warm or
too cold.
Defective test
strip.
Device
malfunction.
EN-17
Read instructions
thoroughly and repeat
the test again.
Shake the control solution
vigorously and repeat the
test again.
Check the expiration date
of the control solution.
Control solution, device,
and test strips should
be at room temperature
(20°C to 25°C / 68°F to
77°F) before testing.
Repeat the test with a
new test strip.
Please contact customer
service.