portable and mobile RF communications equipment and the VRA201.
The VRA201 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the VRA201 can help prevent electromagnetic interferences by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the VRA201 as recommended below, according to the maximum output power of the
communications equipment.
Rated Maximum
output power of
transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply to all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The VRA201 is intended for use in the electromagnetic environment specified below. The customer or the
user of the VRA201 should assure that it is used in such an environment.
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
VRA201 - Instructions for Use - EN
Recommended separation distances between
Separation distance according to frequency of transmitter [m]
d = 1.17 √ ����
150 kHz to 80 MHz
0.12
0.37
1.17
3.70
11.70
IEC 60601 Test level
± 8 kV contact
± 15 kV air
± 2 kV for power
supply lines
100 kHz repetition
frequency
± 1 kV Line-to-line
100 kHz repetition
frequency
± 1 kV Line-to-line
± 2 kV Line-to-ground
0% UT for 0.5 cycle
0 % UT for 1 cycle
and
70% UT
for 25/30 cycles
Single phase: at 0°
d = 1.17 √ ����
80 MHz to 800 MHz
0.12
0.37
1.17
3.70
11.70
Compliance
± 8 kV contact
± 15 kV air
± 2 kV
± 1 kV
± 1 kV
± 2 kV
0% UT for 0.5 cycle
0 % UT for 1 cycle
and
70% UT
for 25/30 cycles
Single phase: at 0°
d = 2.23 √ ����
800 MHz to 2.7 GHz
0.23
0.74
2.33
7.37
23.30
Electromagnetic environment -
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
greater than 30%.
Mains power quality should be
that of a typical commercial or
residential environment.
Mains power quality should be
that of a typical commercial or
residential environment.
Mains power quality should be
that of a typical commercial or
residential environment. If the
user of the VRA201 requires
continued operation during power
mains interruptions, it is
recommended that the VRA201
be powered from an
uninterruptable power supply or
its battery.