Herunterladen Inhalt Inhalt Diese Seite drucken

Ottobock 1C50 Taleo Gebrauchsanweisung Seite 18

Vorschau ausblenden Andere Handbücher für 1C50 Taleo:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
► Remove the adapter cover before the patient leaves the workshop/fit­
ting area.
5.1.1 Applying/removing the footshell
INFORMATION
► Pull the protective sock over the prosthetic foot to prevent noises in the
footshell.
► Always use the prosthetic foot with the footshell.
► Apply or remove the footshell as described in the footshell instructions
for use.
5.1.2 Bench Alignment
INFORMATION
Only use the prosthetic foot with the heel wedge.
TT bench alignment
Required materials: 662M4 goniometer, 743S12 heel height measuring
apparatus, 743A80 50:50 gauge, alignment apparatus (e.g. 743L200
L.A.S.A.R. Assembly or 743A200 PROS.A. Assembly)
Assemble and align the prosthetic components in the alignment apparatus
according to the following specifications:
Heel height: effective heel height (shoe heel height - sole thick­
   
ness of forefoot) + 5 mm
Sagittal alignment: alignment reference line on a–p markings
   
of the footshell (see fig. 2)
Frontal alignment: alignment reference line on m–l markings of
   
the footshell (see fig. 2)
   
Connect the prosthetic foot and prosthetic socket using the selec­
ted adapters. Follow the adapter instructions for use.
   
Sagittal alignment:
Determine the centre of the prosthetic socket with the 50:50
gauge. Align the prosthetic socket centrally to the alignment refer­
ence line.
Socket flexion: individual residual limb flexion + 5°
   
Observe the abduction or adduction position.
18
Bench alignment process

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

1c53 taleo lp