Surface Disinfection With Cleaning; Maintenance; Definition Of Maintenance Concepts - Dräger DigiFlow Gebrauchsanweisung

Elektronischer durchflussmesser
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
EnglishEnglish
Class of
Surface
active
disinfectant
ingredient
Chlorine-
Actichlor plus
releasing
BruTab 6S
agents
Clorox
Professional
Disinfecting
Bleach Cleaner
Dispatch
Hospital
Cleaner
Disinfectant
Towels with
Bleach
Klorsept 17
Oxygen-
Descogen
releasing
Liquid
agents
Descogen
Liquid r.f.u.
Dismozon plus
Dismozon pur
Oxycide
Perform
Virkon
26
Manufacturer
Class of
active
ingredient
Ecolab
Quaternary
ammonium
Brulin
compounds
Clorox
Medentech
Aldehydes
Antiseptica
*
Virucidal against enveloped viruses
Dräger states that oxygen-releasing agents and
chlorine-releasing agents may cause color change in
Bode Chemie
some materials. Color change does not indicate that
the product is not functioning correctly.
Ecolab USA

Surface disinfection with cleaning

Schülke &
Mayr
WARNING
DuPont
Risk due to penetrating liquid
Penetrating liquid may cause the following:
Damage to the device
Electric shock
Device malfunctions
Ensure that no liquid penetrates the device.
1
Remove soiling immediately. Use a cloth
dampened with disinfectant to remove soiling.
2
Perform surface disinfection.
Surface
Manufacturer
disinfectant
Mikrozid
Schülke &
*
sensitive liquid
Mayr
Mikrozid
sensitive
*)
wipes
Mikrozid
alcohol free
*)
liquid
Mikrozid
alcohol free
*)
wipes
*)
acryl-des
Buraton 10 F
Schülke &
Mayr
3
After the product has been exposed to the
disinfectant for the specified contact time,
remove residual disinfectant.
4
Wipe with a cloth dampened with water
(preferably drinking-water quality). Allow the
product to dry.
5
Check the product for visible soiling. Repeat
steps 1 to 5 if necessary.
6
Check the product for visible damage and
replace if necessary.

Maintenance

This chapter describes the maintenance measures
required to maintain the proper functioning of the
medical device. Maintenance measures must be
performed by the personnel responsible.
Do not perform maintenance and preventive
maintenance activities as long as the device is still in
use.

Definition of maintenance concepts

Concept
Definition
Mainte-
All measures (inspection, preventive
nance
maintenance, repair) intended to
maintain and restore the functional
condition of a medical device
Inspection
Measures intended to determine and
assess the actual state of a medical
device
Preventive
Recurrent specified measures in-
mainte-
tended to maintain the functional
nance
condition of a medical device
Instructions for use DigiFlow