Herunterladen Inhalt Inhalt Diese Seite drucken

TESTBOY TV 465 PRO Bedienungsanleitung Seite 82

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

®
Testboy
TV 465
Memory..............................................1500 memory locations
The error in operating conditions could be at most the error for reference conditions
(specified in the manual for each function) +1 % of measured value + 1 digit, unless
otherwise specified in the manual for particular function.
Communication transfer speed
RS232 interface ................................ 9600 bps, 1 start bit, 8 data bits, 1 stop bit
RS232 connector .............................. PS/2 connector, female
USB interface .................................... 115200 bps
USB connector .................................. type B
Protection pre-tests
External voltage between LN and PE (DC and AC).
Excessive leakage between S/EB1 and PE (DC and AC).
5.
Connectivity (fuse) pre-test
Appliance not switched on or too high resistance between L and N
Maximum resistance for connectivity pre-test .................................. 30 k