Device Use and Control with High Flow Operating Mode
EN
26
6.6.3

Stopping Insufflation

1. Press the [STOP] function field. The following values are displayed:
• Gas consumption display: last measured value.
• Actual pressure: Current measured value
• Actual gas flow: 0 l/min
• Nominal pressure: last set value In case of exceeding the safety limit, the
nominal pressure value will be reset to the safety limit value.
• Nominal gas flow: last set value
The display status field indicates Insufflation Stopped.
2. Remove the tube set from the device. Observe applicable hygiene regulations
when disposing of the tube set.
CAUTION!
If the tube set remains connected to the device, there is the danger that leftover
fluid in the tube or the instruments will be sucked into the device.
3. Close the gas supply.
4. Use the ON/OFF switch to turn the device off.
NOTE!
Observe applicable hygiene regulations when disposing of the tube set.