28
Assignment for plug 1:
Supply voltage and frequency output signal.
Connect
Device
connector
6
1
5
7
2
3
4
Top view
1)
Complementary signals RS422; with cable lengths exceeding 10m, we recommend
using a termination resistor R = 120 Ohm between the wires (wh) and (rd).
2)
RS422: Pin 1 corresponds to A, Pin 4 corresponds to B.
NOTE
These torque flanges are only intended for operation with a DC supply
voltage. They must not be connected to older HBM amplifiers with
square‐wave excitation. This could destroy the connection board resistances
or cause other faults in the amplifiers.
HBM
Assignment
or
pin
Torque measurement signal
1
(frequency output; 5 V
2
Supply voltage 0 V;
Supply voltage 18 V  30 V
3
Torque measurement signal
4
(frequency output; 5 V
5
Measurement signal 0 V;
Shunt signal trigger 5 V  30 V
6
Shunt signal 0 V;
7
Shielding connected to enclosure ground
1),2)
)
1),2)
)
symmetrical
T40FM
Wire
Sub‐D
color
plug
Pin
wh
13
bk
5
bu
6
rd
12
gy
8
gn
14
gy
8
A3276-2.1 en/de