Front / Remote Selection

By pressing CYCLE NUMBER and PUMP CURRENT pushbuttons
together for at least 2 seconds, the processor enters in a routine
where it is possible to program the controller. In this routine, the
CYCLE pushbutton is used for choosing/changing the value or
condition; the PUMP CURRENT pushbutton is used to enter and
confirm the value.
The display shows:
S
S
where: XXX = YES or NO.
If YES is selected, the Soft Start mode allows the pump to ramp-
up the Normal speed within ten steps.
When NO is selected, the Soft Start mode is deselected and the
ramp-up of the pump will be done within 75 seconds.
The controller is factory set to YES.
The Soft Start mode may be deselected/selected only when the pump is
NOTE
stopped.
If necessary press CYCLE NUMBER to select YES or NO, then
enter the selection by pressing the PUMP CURRENT pushbutton.
Turbo-V 70 PCB Controller User Manual / 87-900-871-01 (E)