13 Error Messages and Warnings

Error message
Out of Specification
Out of Range
Temperature Range
Density Range
Criterion
Low Battery
Battery Empty
Error 01
Error 02
Error 03
Error 04
NOTICE
If the error messages "Error 01" to "Error 04" occur, send in your
instrument for service.
C96IB01
Cause
The measured sample temperature is outside the
specifications.
The measured value is outside the specified range.
Possible reasons:
The temperature is too high or too low.
The density is too high or too low.
The measured value is invalid.
The temperature of the reference liquid for
readjustment is outside the temperature
specifications (+15 to +25 °C).
The density of deionized water at 20 °C measured
during readjustment is outside the allowed limits
(tolerance 0.01 g/cm
The density of deionized water at 20 °C measured
during readjustment deviates from factory
adjustment by more than 0.01 g/cm
The battery is almost empty.
The battery is empty.
The temperature is invalid.
The period is invalid.
DMA 35 has no adjustment data.
Calculation error.
3
).
3
.
EN
65