9.7
Performing a Measurement
1.
Select the measuring unit (see Chapter 9.1) or measuring method (see Chapter
9.2).
2.
Define a sample ID if you want to assign the measurement results to a specific
sample (see Chapter 9.3).
3.
Fill the measuring cell with the sample.
4.
Press the
The measured value is displayed as soon as the result has stabilized. The
measured value with all corresponding data is now saved in the results list.
5.
Press <OK> to return to the measuring mode.
9.7.1 Deleting the Measured Value
•
Press
display.
The measured value is deleted and the instrument returns to the measuring
mode.
9.7.2 Printing the Measured Value
You can transfer the measured value to a printer via the IrDA interface. For a
compatible printer, contact your Anton Paar representative.
1.
Allocate the softkey the command <Print> (see Chapter 8.7).
2.
Switch on the printer with the IrDA interface.
3.
Press the <Print> key when the measured value appears on the display.
4.
Hold the IrDA interface of DMA 35 to the IrDA interface of the printer to establish
communication between the printer and density meter.
The measured value is printed.
You can also print out all or only the last measuring data (see Chapter 9.10).
9.7.3 Emptying the Measuring Cell
1.
Place the filling tube in a vessel which is suitable for sample disposal.
2.
Empty the measuring cell by pressing the pump lever.
C96IB01
key on the back of DMA 35.
instead of <OK> when the measured value appears on the
EN
49