2.
Empty the measuring cell by pressing the pump lever.
3.
Open the pump lock and remove the pump.
4.
Hold the pump housing under running water and rinse it thoroughly.
5.
Insert the pump and close the pump lock.
6.
Empty the measuring cell by pressing the pump lever.
10.4 Cleaning the Housing and Display
To clean the housing and display, use a soft cloth dipped in ethanol or warm water. If
necessary, you can use a mild solvent (pH < 10).
10.5 Storing the Instrument
Before storing the instrument for a longer period of time, clean the measuring cell as
described in Chapter 10.1. This prevents any liquid residues freezing or drying in the
measuring cell, which could destroy the cell or lead to measuring inaccuracies.
For storage lasting less than one day, fill the measuring cell with deionized water or
solvent. If you have filled the liquid via the plastic syringe, leave the syringe in the
adapter to stop the liquid running out.