APPENDIX A Specifications

Notes:
● Specifications may be changed without prior notice.
● The circuit diagrams, the list of components, the illustration of diagrams, and the detailed
rules of calibration, are provided exclusively to professional personnel authorized by our company.
● The equipment has been calibrated , users do not to calibrate. In order to ensure the
accuracy of the probe, please change the probe once a year. Make sure that the type of
probe need to be specified.
Specifications:
Display:
Data: SpO
%, PR, PI, pulse bar
2
Others: connection status of probe and other alarm information.
Equipment data update period
As shown in the following figure. Data update period of slower average is 8s.
Alarm:
Alarm: SpO
and pulse rate value, probe off, battery exhausted
2
Alarm mode: audible alarm, visual alarm and information
Alarm limits range: SpO
70%~100%, PR 30bpm-250bpm
2
Default limits: SpO
High 100%, low 90%; PR High 100 bpm, low 60 bpm
2
SpO
2
Measurement range: 70% ~ 100%
Resolution: 1%
Accuracy: 70% ~ 100%: ±2%;<70% unspecified
Pulse Rate
Measurement range: 30bpm ~ 250bpm
Resolution: 1 bpm
Accuracy: ±2 bpm or 2%(The larger)
Perfusion Index
Measurement range: 0.1% ~ 20%
Accuracy: 0.1%~1.0% ±0.2%; 1.1%~20% ±20%
PULSE OXIMETER
- 26 -
INSTRUCTION MANUAL
Probe
Emitter: OL660905HM2-2(H2)-C
Receiver: OP30TMF-3
Probe LED Specifications:
Wavelength
RED
660±3nm
IR
905±10nm
Environment Requirements
Operation temperature: 0℃ ~ 40℃
Operation humidity: ≤80%,no condensation
Storage / transport temperature: -20℃ ~ +70℃
Storage / transport humidity: ≤93%, no condensation
Power supply: Three AA alkaline batteries or adapter
Working time: alkaline batteries: more than 10 hours
Atmosphere pressure: 86kPa~106kPa
AC adapter(optional)
Input Voltage: AC 100V ~ 240V
Input Frequency: 50Hz ~ 60Hz
Output Voltage: DC 5V±5%
Output Current: 2A MAX
Fuse
Type: 1206L050
I(hold)0.5A, I(trip)1A, V(max)15V
Store and Replay
Store and replay 72 hours SpO
% and Pulse rate value, the time interval is 4 seconds.
2
Outline of Product
Dimension: 125mmX60mmX30mm
Weight: 195g (excluding the batteries)
Equipment Classification
Classification according to IEC-60601-1
According to the type of protection against
Electrical shock
According to the degree of protection against
Electrical shock
According to the degree of protection against
harmful ingress of water.
According to the methods of sterilization or
disinfection.
MD300M
Radiant Power
3.2mW
2.4mW
Internal electrical power source equipment
and Class II equipment
Type BF equipment