CleanSpace Respirators
Name
Mask Size
Neck Support
Date of Quantitative Fit Test
Fit Factor on Test
IF A SATISFACTORY FIT FACTOR CANNOT BE ACHIEVED WITH ANY OF THE MASKS, CLEANSPACE RESPIRATOR MUST NOT BE
USING CLEANSPACE RESPIRATOR WITH THE HALF MASK.
Complete the following steps each time you use CleanSpace Respirator
STEP 1 – INSPECT THE EQUIPMENT
Before each entry into a contaminated area, the following inspections must be performed:
Visually check the entire CleanSpace Respirator system including the blower, mask, harness, filter adaptor and filter. If
parts are missing or damaged replace them only with approved CleanSpace Respirator parts before proceeding. Check
the top and bottom case of the blower, and the Filter Adaptor (if fitted) for cracks or other damage. Do not use the
device if there is any damage.
Remove the filter and check it carefully. The seal must be clean and free from damage of any kind. If necessary it can be
cleaned with a cloth dampened with water. Examine the surfaces behind the filter carefully for any sign that dust has
leaked past a damaged seal. If found, replace the filter. The body of the filter must not be cracked or show any sign of
damage. Examine the filter carefully for any sign that it has sustained an impact or been scratched. If any sign of impact
or scratching is found, discard the filter. At the end of your inspection, refit a filter and check that it is secure (see
Changing the Filter, below)
WARNING: NEVER CLEAN THE FILTER WITH COMPRESSED AIR OR A BRUSH. HEPA FILTERS ARE EASILY DAMAGED BY THE USE
Misuse of CleanSpace Respirator by cleaning the filter may result in overexposure to contaminants and lead to sickness or
Check that the battery is fully charged by pressing the power button. All three battery LEDs must light. Check that the
battery has enough charge to complete the work period.
Check both bellows for splits or holes. Check that the bellows have not become distorted so as to partial ly or fully close
the air path to the mask.
Check the mask to ensure that there are no cracks, tears or dirt; check the mask is not distorted.
Check the exhalation valve for damage or dirt build up. If it is dirty, remove the exhalation valve cover. Remove any dirt,
hairs or anything that could affect the seal of the valve against its seat. Check that the valve seat is clean. Reinstall the
valve cover. If the valve is damaged, replace the mask.
Check the harness is intact. It must be adjusted to support some of the weight of the CleanSpace Respirator.
STEP 2 – CALIBRATE THE POWER UNIT
CleanSpace Respirator contains a system that detects and corrects for changes in altitude. This system requires re-calibration
whenever it experiences a change in temperature of more than 20°C. To re-calibrate:
1.
With CleanSpace Respirator in standby mode (one or more green LEDs lit), press and hold both the Power button and the
FLOW TEST button. If no LEDs are lit, press the Power button once to enter standby mode.
2.
When both the blue and red LEDs light, release both buttons.
3.
Hold the CleanSpace Respirator still until the red and blue LEDs turn off (10 – 15 seconds)
4.
Calibration is complete.
Small / Large
USED
OF COMPRESSED AIR OR BY BRUSHING.
death.
9