2. DISASSEMBLY OF THE NITROGEN/AIR SUPPLY

Before disassembly the nitrogen/air supply, no nitrogen/air pressure must remain in the console.
Steps
What to do
IMPORTANT WARNINGS
1
Close the gas tank.
After each surgical session, the console must be disconnected
from the dry medical grade nitrogen gas supply, or dry and filtered
medical grade air supply (internal circuit or tank).
For safety reasons, no nitrogen or air pressure must remain in the
console.
2
Press the vacuum
footpedal.
3
Operate the turbine by
pressing the Forward/
Turbine footpedal for
few seconds until the
turbine runs out of gas
and comes to a stop.
4
Stop the vacuum.
5
Switch OFF the console.
6
Press the Residual Gas
Drain Switch until there
is no pressure left in the
console
7
Disconnect the nitrogen/
Disconnect the gas supply hose:
air gas supply (internal
• first push the blue ring toward the unit (step 1),
circuit or tank) from the
• then pull the hose towards the back (step 2)
control unit.
H. AUTOMATED MODE (M2, M2SU, ONE USE-PLUS, EPI-K™)
Steps
What to do
IMPORTANT WARNINGS
1
Select the automated
Press « Mode » button (A13) until green LED (A14) is ON
mode
according the selected microkeratome.
« OP » appears on the display.
2
Proceed a vacuum test
Refer to chapter VIII.C
3
Plug the motor system
To plug One Use-Plus handpieces: A19
into the control unit
To plug M2-M2SU handpiece: A21
To plug Epi-K
4
Select the advance
The advance speed 1 is slower than the advance speed 2. The
speed (for M2, M2SU
blade oscillation rate is not modified.
and One Use-Plus)
See applicable nomograms:
• M2-M2SU : #65023
• One Use-Plus with plastic ring : #65041
• One Use-Plus with metal ring: #65067, 65077EN
• One Use-Plus for artificial chamber : #65101
5
Control unit is ready
Refer to user manual:
to use.
• M2-M2SU: #65013
• One Use-Plus with plastic ring: #65039
• One Use-Plus with metal ring: #65040
• One Use-Plus for artificial chamber: #65009
• Epi-K™ with metal ring: #65043
• Epi-K™ with plastic ring: #65044
Always check intraocular pressure with a tonometer (#19042) prior
to using any keratome.
handpiece: A22
TM
IX. USE OF FOOTPEDAL
A. COMPATIBILITY FOOTPEDAL-CONSOLE
Related
picture
Footpedals (#19361 and 19381) with serial number inferior to 7000 are not compatible with
EVOLUTION 3E consoles (#19380) having a serial number superior to 7000.
EVOLUTION 3E consoles (#19380) with serial number superior to 7000 are compatible only with
footpedals (#19361 and 19381) having a serial number superior to 7000.
C1
Footpedal
(#19361 and 19381)
C2
5000 < sn
5000 < sn < 7000
sn > 7000
B. COMPATIBILITY FOOTPEDAL-HANDPIECE
C1
B9
Footpedal #19361 can be used with M2 and One Use-Plus handpieces, as well with turbines for
B4
ONE and CB. Footpedal #19361 can not be used with the Epi-K™ handpiece.
Footpedal #19381 can be used with M2, One Use-Plus and Epi-K™ handpieces, as well with
turbines for ONE and CB.
B2–B3
C. MANUAL MODE
D3a-D3b
What to do
Press the vacuum footswitch (C1)
Related
once: vacuum is ON
Press again the vacuum footswitch
picture
(C1): vacuum is OFF
A13
A14
Press the forward footswitch (C2):
A15
blade oscillation
A8
A19
A21
Press the backward footswitch (C3):
A22
no action
A20
EVOLUTION 3E console (#19380)
5000 < sn < 7000
compatible
compatible
not compatible
IMPORTANT WARNINGS
The system displays the actual vacuum value in mmHg (A7).
When a suction ring is fixated on the eye, the vacuum value
displayed must be less than 250 mmHg.
If the value displayed is above 250 mmHg, please refer to chapter
X.
In case vacuum is not activated, the forward footswitch does not
work.
Blade oscillation stops once forward footswitch is not pressed any
longer.
Do not press footpedal (C2) and (C3) at the same time.
#65060/INTL-I-06.2017
sn > 7000
not compatible
not compatible
compatible
Related
picture
C1
C2
C3
7
MORIA

Werbung