Herunterladen Inhalt Inhalt Diese Seite drucken

Ambient Conditions; Electrical Specifications - Benning CM 1-4 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6.

Ambient conditions

-
The BENNING CM 1-4 is intended for measurements under dry ambient
conditions
-
Maximum barometric height for measurements: 2000 m
- Overvoltage category / installation category: IEC 60664/ IEC 61010 → 600 V
category IV; 1000 V category III
-
Contamination class: 2
-
Protection category: IP 65 (DIN VDE 0470-1 IEC/ EN 60529)
6 - first index: protection against access to dangerous parts and protection
against solid impurities, dustproof
5 - second index: protected against water jets. The device can also be used
in the rain.
-
Operating temperature and relative air humidity:
For operating temperatures from - 15 °C to 30 °C: relative air humidity lower
than 80 %.
For operating temperatures from 31 °C to 40 °C: relative air humidity lower
than 75 %.
For operating temperatures from 41 °C to 55 °C: relative air humidity lower
than 45 %.
-
Storage temperature:
The BENNING CM 1-4 can be stored at temperatures between - 20 °C and
+ 60 °C (air humidity 0 to 80 %). During storage, the battery should be re-
moved.
7.

Electrical specifications

Note: The measuring accuracy is specified as the sum of:
-
a relative part of the measured value and
-
a number of digits (i.e. counting steps of the last digit).
This measuring accuracy applies to temperatures from 23 °C ± 5 °C and a rela-
tive air humidity lower than 80 %.
The measuring value is gained and indicated as effective value (True RMS, AC
coupling). In case of non-sinusoidal curves, the indicating value becomes inac-
curate. Thus, an additional error occurs for the following crest factors:
crest factor from 1.0 to 2.0 additional error + 1.0 %
crest factor from 2.0 to 2.5 additional error + 2.5 %
crest factor from 2.5 to 3.0 additional error + 4.0 %
max. crest faktor:
crest faktor 3 @ 5000 Digit
crest faktor 1.5 @ 9999 Digit
7.1 AC voltage ranges
Overload protection: 1000 V
Function
With batteries
Without batteries
The input resistance for voltage measurement depends on the applied voltage:
approx. 20 kΩ at 50 V to approx. 305 kΩ at 1000 V
U > 300 V: tON: 30 s, tOFF: 240 s
For frequencies higher than 65 Hz, the lower measuring range is 8 V.
*1
Valid for temperatures < 35 °C ... > - 15 °C
*2
7.2 DC voltage ranges
Overload protection: 1000 V
Function
With batteries
Without batteries
The input resistance for voltage measurement depends on the applied voltage:
approx. 20 kΩ at 50 V to approx. 305 kΩ at 1000 V
U > 300 V: tON: 30 s, tOFF: 240 s
Valid for temperatures < 35 °C ... > - 15 °C
*2
09/ 2019
AC/DC
Measuring range
6.0 V - 999.9 V
*1
45.0 V - 999.9 V
*2
AC/DC
Measuring range
7.0 V - 999.9 V
*1
35.0 V - 999.9 V
*2
BENNING CM 1–4
Resolution
in frequency range 45 Hz - 400 Hz
0.1 V
0.1 V
Resolution
0.1 V
0.1 V
Measurement accuracy
± (1.5 % of reading + 7 digits)
Measurement accuracy
± (1.0 % of reading + 4 digits)
24

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis