6.

Ambient conditions

-
The BENNING MM P3 is intended for measurements under dry ambient condi-
tions,
-
Maximum barometric height for measurements: 2000 m,
-
Overvoltage category / installation category: IEC 60664-1/ IEC 61010-1 → 300 V
category III; 600 V category II,
-
Contamination class: 2,
-
Protection category: IP 30 DIN VDE 0470-1 IEC/EN 60529
IP 30 means: Protection against access to dangerous parts and protection
against solid impurities of a diameter > 2.5 mm, (3 – first index).
No protection against water, (0 – second index).
-
Operating temperature and relative air humidity:
For operating temperatures from 0 °C to 30 °C: relative air humidity less than 80 %,
For operating temperatures from 31 °C to 40 °C: relative air humidity less than 75 %,
For operating temperatures from 41 °C to 50 °C: relative air humidity less than 45 %,
-
Storage temperature: The BENNING MM P3 can be stored at temperatures be-
tween - 20 °C and + 60 °C (air humidity of 0 to 80 %). During storage, the bat-
teries should be removed.
7.
Electrical specifications
Note: The measuring accuracy is specified as the sum of:
-
a relative part of the measured value and
-
a number of digits (i.e. counting steps of the last digit).
This measuring accuracy applies for temperatures from 18 °C to 28 °C and a relative
air humidity less than 80 %.
7.1 DC voltage ranges (Switch setting: V
The input resistance is 10 MΩ.
Measuring range
400 mV
4 V
40 V
400 V
600 V
7.2 AC voltage ranges (Switch setting: V
The input resistance is 10 MΩ in parallel 100 pF.
Measuring range
400 mV
4 V
02/ 2011
Resolution
0.1 mV
± (0.7 % of the measured value + 5 digits)
1 mV
± (0.6 % of the measured value + 2 digits)
10 mV
± (0.6 % of the measured value + 2 digits)
100 mV
± (0.6 % of the measured value + 2 digits)
1 V
± (0.7 % of the measured value + 5 digits)
Resolution
0.1 mV
± (1.5 % of the measured value + 5 digits)
1 mV
± (0.9 % of the measured value + 5 digits)
BENNING MM P3
mV
)
DC
DC
Measuring accuracy
, Hz, %)
AC
Measuring accuracy *
frequency range 50 Hz - 500 Hz
within the
1
Overload
protection
600 V
DC
600 V
DC
600 V
DC
600 V
DC
600 V
DC
Overload
protection
600 V
eff
600 V
eff
16