3. Specifcations
3. Specifcations
Reference pressure sensors
Pressure range
Accuracy
Type of pressure
Pressure units
Base instrument
Instrument
Instrument version
Dimensions
Weight
Ingress protection
Pressure generation
Display
Screen
Resolution
Keyboard
Menu language
Connections
Pressure connections
Wetted parts
Voltage supply
Power supply
Power consumption
Battery type
Battery life
Permissible ambient conditions
Media
Operating temperature
Storage temperature
Relative humidity
WIKA operating instructions, model CPC2000
mbar
1
10
% FS
0.3
0.1
Positive or negative gauge pressure or diferential pressure
Pa, kPa, hPa, bar, mbar, psi, inH
desktop instrument with carry handle
102.6 x 257 x 271 mm without carry handle
4.6 kg
IP 20
internal, electric pump
backlit, multiline graphic display
5 (4 digits for 50/500 mbar measuring range)
Membrane keypad
German, English, Spanish and French
6.6 x 11 (hose diameter, D = 6 mm)
Ni, Al, CuBe, PU
DC 24 V, 1 A
24 VA
Li-Ion
approx. 8 h
Ambient air
10 ... 40 °C
-10 ... +70 °C
30 ... 80 % r. h. (non-condensing)
50
100
500
0.1
0.1
0.1
O, inHg, mmHg, Torr
2
GB
1,000
0.1
9