8. Maintenance, cleaning and service (recalibration)

8. Maintenance, cleaning and service (recalibration)
8.1 Maintenance
This low-pressure controller is maintenance-free.
Repairs must only be carried out by the manufacturer.
The safety of the operator may be endangered by the low-pressure controller if, for
example:
there is visible damage to the instrument.
the instrument is not working as specifed.
the instrument has been stored under unsuitable conditions for an extended period of
time.
If there is any doubt, please return the low-pressure controller to the manufacturer for
repair or maintenance.
8.2 Cleaning
CAUTION!
Before cleaning, correctly disconnect the low-pressure controller from
the pressure supply, switch it of and disconnect it from the mains.
Clean the instrument with a moist cloth.
Electrical connections must not come into contact with moisture.
Wash or clean the dismounted instrument before returning it, in order to
protect persons and the environment from exposure to residual media.
Residual media on the dismounted low-pressure controller can result
in a risk to persons, the environment and equipment. Take sufcient
precautionary measures.
For information on returning the instrument see chapter 10.2 "Return".
8.3 Recalibration
DKD/DAkkS certifcate - Certifcates:
We recommend that the instrument is regularly recalibrated by the manufacturer, with
time intervals of approx. 12 months. Every factory recalibration includes, additionally, an
extensive free-of-charge check of all system parameters with respect to their compliance
with the specifcation. The basic settings will be corrected if necessary.
WIKA operating instructions, model CPC2000