4. Design and function
4.6 Pressure connections
Diferential pressure instruments are connected to the
■
instrument.
Relative pressure instruments are connected to the
■
open to atmosphere).
Instruments for negative gauge pressure are connected to the
■
(
connection is open to atmosphere).
j
Only connect test and calibration installations once the system has been
depressurised!
Never perform functional tests with compressed or medical air. This can
damage instruments with low pressure ranges.
The connection to the low-pressure controller is made via a 6.6 x 11 mm pressure connec-
tion. The hose diameter should be 6 mm.
4.7 Overpressure protection
The maximum permissible pressure must not be exceeded:
Pressure range
1 mbar
10 mbar
50 mbar
100 mbar
500 mbar
1,000 mbar
WIKA operating instructions, model CPC2000
j
connection (
j
Oberload
5 times overload
5 times overload
5 times overload
5 times overload
2 times overload
2 times overload
and the
connections of the
i
connection is
i
connection
i