► Do not grind the prosthetic foot or footshell.
5.1 Alignment
INFORMATION
There is a plastic adapter cover on the proximal connection of the prosthet­
ic foot. This protects the connecting section from scratches during the
alignment and trial fitting.
► Remove the adapter cover before the patient leaves the workshop/fit­
ting area.
5.1.1 Applying/removing the footshell
Inserting the foot into the footshell
Recommended tools: 2C100 changing device for footshell
>
1) Pull the Spectra sock over the prosthetic foot, avoiding any wrinkles.
2) Slide the prosthetic foot into the footshell. Insert the carbon spring into
the slot on the toe insert (see fig. 4).
3) Push the prosthetic foot into the heel area of the footshell until it locks
into place (see fig. 6).
4) Place the connection cap on the footshell (see fig. 7).
Removing the footshell
Recommended tools: 2C100 changing device for footshell
>
1) Push the heel area of the footshell to the rear and pull the prosthetic foot
up.
2) Remove the prosthetic foot from the footshell.
5.1.2 Bench Alignment
TT bench alignment
Recommended tools:
662M4 goniometer, 743S12 heel height measuring apparatus,
743A80 50:50 gauge, alignment apparatus (e.g. 743L200
L.A.S.A.R. assembly or 743A200 PROS.A. assembly)
Perform the alignment and assembly of the prosthetic components in the
alignment apparatus according to the following specifications:
   
Heel height:
Effective heel height (shoe heel height – sole thickness of fore­
foot) + 5 mm
16 | Ottobock
Bench alignment process