DK50-10
5.1.
Free air delivery (FAD) correction due to elevation
FAD correction table
Elevation [mamsl]
FAD [l/min]
FAD output related to conditions:
Elevation: 0 MASL
Atmospheric pressure : 101325 Pa

6. PRODUCT DESCRIPTION

6.1.
Model variations and their uses
The compressor is used as a source of clean, oil-free compressed air to power dental equipment,
instruments and devices in laboratories where the parameters and properties of the compressed air are
suitable to the specific application.
Compressed air supplied by the compressor is unsuitable for use with artificial lung
ventilation devices without further filtration
Any use of the equipment outside the framework defined herein is considered inappropriate use. The
manufacturer is not liable for any damages or injury occurring as a result of inappropriate use.
6.2.
Description
Compressor models are designed for the following uses:
DK50-10 Z compressor - a base-mounted compressor for stand-alone installation in a given space
DK50-10 Z/K compressor - a base-mounted compressor with a condensate and filtration unit (KJF1)
DK50-10 Z/M compressor - a base-mounted compressor with an air dryer
DK50-10 S compressor - an enclosed compressor with effective noise dampening for in-office installations
DK50-10 S/K compressor - an enclosed compressor with a condensate and filtration unit (KJF1)
DK50-10 S/M compressor - an enclosed compressor with an air dryer
DK50-10 Z
08/2017
0 - 1500
1501 - 2500
FAD x 1
FAD x 0.8
Temperature : 20°C
Relative humidity : 0%
DK50-10 Z/M
2501 - 3500 3501 - 4500
FAD x 0.71
FAD x 0.60
DK50-10 S
DK50-10 S/M
- 9 -