Do not insert or withdraw the guidewire/stylet forcibly from any
component. The wire may break or unravel. If the guidewire
becomes damaged, the guidewire along with the introducer needle,
Vascu-Sheath
together.
Federal Law (USA) restricts the device to sale by or on the order of a
physician.
This catheter is for Single Use Only.
Do not re-sterilize the catheter or accessories by any method.
Re-Use may lead to infection or illness/injury.
The manufacturer shall not be liable for any damages caused by
reuse or re-sterilization of this catheter or accessories.
Contents sterile and non-pyrogenic in unopened, undamaged
package.
Do not use catheter or accessories if package is opened or damaged.
Do not use catheter or accessories if any sign of product damage is
visible.
CATHETER PRECAUTIONS:
Use only Medcomp
Do not use sharp instruments near the extension tubing or catheter
lumen.
Do not use scissors to remove dressing.
Catheter will be damaged if clamps other than what is provided with
this kit are used.
Clamping of the tubing repeatedly in the same location may weaken
tubing. Avoid clamping near the luers and hub of the catheter.
Examine catheter lumen and extensions before and after each
treatment for damage.
Tape injection caps between treatments to prevent accidental
removal.
Do not clamp arterial extension when stylet is in catheter.
To prevent accidents, assure the security of all caps and bloodline
connections prior to and between treatments.
Use only Luer Lock (threaded) Connectors with this catheter.
Repeated over tightening of bloodlines, syringes, and caps will
reduce connector life and could lead to potential connector failure.
Always tape luer locks to bloodlines during treatment to safeguard
against accidental disconnection.
INSERTION SITES:
Warning: Physician discretion is strongly advised when inserting this
catheter in patients who are unable to take or hold a deep breath.
The patient should be in a modified Trendelenburg position, with
the upper chest exposed and the head turned slightly to the side
opposite the insertion area. A small rolled towel may be inserted
between the shoulder blades to facilitate the extension of the chest
area.
introducer, dilator, or stylet must be removed
®
STERILIZED BY ETHYLENE OXIDE
STERILE EO
Lock Right
®
2
Adapters with this catheter.
®