Do not insert or withdraw the guidewire/stylet forcibly from any
component. The wire may break or unravel. If the guidewire
becomes damaged, the guidewire along with the introducer needle,
Vascu-Sheath
together.
•
Federal Law (USA) restricts the device to sale by or on the order of a
physician.
•
This catheter is for Single Use Only.
•
Do not re-sterilize the catheter or accessories by any method.
•
Re-Use may lead to infection or illness/injury.
•
The manufacturer shall not be liable for any damages caused by
reuse or re-sterilization of this catheter or accessories.
•
Contents sterile and non-pyrogenic in unopened, undamaged
package.
•
Do not use catheter or accessories if package is opened or damaged.
•
Do not use catheter or accessories if any sign of product damage is
visible.
CATHETER PRECAUTIONS:
•
Use only Medcomp
•
Do not use sharp instruments near the extension tubing or catheter
lumen.
•
Do not use scissors to remove dressing.
•
Catheter will be damaged if clamps other than what is provided with
this kit are used.
•
Clamping of the tubing repeatedly in the same location may weaken
tubing. Avoid clamping near the luers and hub of the catheter.
•
Examine catheter lumen and extensions before and after each
treatment for damage.
•
Tape injection caps between treatments to prevent accidental
removal.
•
Do not clamp arterial extension when stylet is in catheter.
•
To prevent accidents, assure the security of all caps and bloodline
connections prior to and between treatments.
•
Use only Luer Lock (threaded) Connectors with this catheter.
•
Repeated over tightening of bloodlines, syringes, and caps will
reduce connector life and could lead to potential connector failure.
•
Always tape luer locks to bloodlines during treatment to safeguard
against accidental disconnection.
INSERTION SITES:
Warning: Physician discretion is strongly advised when inserting this
catheter in patients who are unable to take or hold a deep breath.
•
The patient should be in a modified Trendelenburg position, with
the upper chest exposed and the head turned slightly to the side
opposite the insertion area. A small rolled towel may be inserted
between the shoulder blades to facilitate the extension of the chest
area.
introducer, dilator, or stylet must be removed
®
STERILIZED BY ETHYLENE OXIDE
STERILE EO
Lock Right
®
2
Adapters with this catheter.
®
-2-