Technical Information

Connection of the High Vacuum Flange

Connection of the High Vacuum Flange

For ISO-K flange connections, fix the two flanges with the clamps as
shown in the figure.
Figure 30
WARNING!
It is mandatory to connect the pump in such a way to withstand the torque
specified in the "IN-STRUCTION FOR USE" section. Specifically pay attention
to clamp design, material of clamps and bolts and bolt fixing torque.
Exercise care when tightening nuts and bolts to avoid creating dents in the
CAUTION!
envelope as this may cause the pump rotor to lock.